有MD,F(xiàn)igure 13其實就變成了接觸電流斷開接地測量金屬可觸及部件的情況。請看下圖:
標準原文:
In the SINGLE FAULT CONDITION of interruption of any one PROTECTIVE EARTH CONDUCTOR (when applicable, see 8.1 b)), measure with MD between earth and any part of the ENCLOSURE(S) that is normally PROTECTIVELY EARTHED.
在斷開任意一根保護接地導(dǎo)線的單一故障狀態(tài)下,用MD在地和正常情況下保護接地的外殼任意部分之間測量。
The EARTH LEAKAGE CURRENT flowing through the PROTECTIVE EARTH CONDUCTOR is not a HAZARD per se. The PATIENT and OPERATOR are protected by specifying appropriately low values for PATIENT LEAKAGE CURRENT and TOUCH CURRENT in NORMAL CONDITION and in relevant SINGLE FAULT CONDITIONS including interruption of the PROTECTIVE EARTH CONDUCTOR. However, an excessive EARTH LEAKAGE CURRENT could pose a possible problem for the installation’s earthing system and any circuit breakers operated by current imbalance
detectors.
See also IEC 60364-7-710 [10].
流過保護接地導(dǎo)線的對地漏電流就其本身而言沒有危險(源)。通過對正常狀態(tài)下與包括斷開保護接地導(dǎo)線在內(nèi)的相關(guān)單一故障狀態(tài)下的患者漏電流和接觸電流規(guī)定合適的低值來保護患者和操作者。然而,過大的對地漏電流可能使設(shè)施的接地系統(tǒng)產(chǎn)生問題,而且電流不平衡檢測器會引發(fā)任何斷路器動作。
參見IEC 60364-7-710 [10]。
既然跟人體安全無關(guān),那要MD干啥?那這個測試的目的是干嘛?更多的是考慮漏電保護吧?(防止接地故障電流漏電保護器動作導(dǎo)致斷電,這個故障,對于醫(yī)院來說還是十分危險的情況)。我覺得這個很像IT里面的保護導(dǎo)體電流的測量。
說了那么多,其實搞清楚儀表怎么測試的就好了。關(guān)于實際測量電路,我仍在等廠商確認。不過可以先參考說明書,先了解了解保護導(dǎo)體電流。
接觸過的漏電流測試裝置不多,有菊水TOS 3200、民勝、日置ST5540等,大概翻閱了一下說明書。
看TOS 3200:
只關(guān)注區(qū)別,不要在意他能不能測60601-1。保護導(dǎo)體電流測試電路不涉及網(wǎng)絡(luò),里面串聯(lián)了一個RES阻抗。這邊的阻抗值是:XXXΩ(目前跟上海菊水溝通過,說是這邊給用戶自己外接的,what???等日本菊水解釋。),原理很簡單,RES足夠小,測出電壓就,電流就越準。至于這邊為啥不用更小的電阻,可能跟電壓測量儀表的精確度有關(guān),單獨串一個電流表也是可以的。
在實際測試對地漏電流過程中,儀器或者設(shè)備并沒有加入人體模擬網(wǎng)絡(luò),這也好解釋為啥對地漏電流正常模式是5mA(串的是RES,很小)限值,接觸電流斷開接地測量金屬可觸及部件限值是500μA(串的是網(wǎng)絡(luò))了。所以依舊堅持自己的觀點:Figure 13放一個MD有點不妥。
時間比較趕,羅里吧嗦說了這些,邏輯都有點問題,我干脆把圖畫出來了,如下: